Composition
Two to six exposures from one acquisition are combined into a derived DICOM image. The image is stored with the source images and does not replace them. One exposure, the reference tile, is copied without change, including its display window. The remaining exposures are mapped onto that grey scale from the anatomy shared in the overlap. The order of grey values is preserved: air, then soft tissue, then bone. The mapping does not add, remove, or locally alter structure.
Review
Each result is marked PASS, REVIEW, or FAIL, and includes a confidence score and the reasons for that mark. The seam is indicated, and a QC image shows the overlap. REVIEW and FAIL are not transferred to PACS until a user confirms them. The decision is applied on the server.
Record
The same input and the same configuration produce the same composite. The measurement scale is written in the DICOM header, including the reference plane and the source of that scale. Angles do not depend on the plane height. Absolute lengths do.